Climate chamber ICHeco

Climate chamber, ICH eco, 108L to 749L

$0.00

#Climate Chamber

#Constant Climate Chamber

#Climate Chamber ICH eco

Model: #ICH110 #ICH110C #ICH110L #ICH110eco #ICH110eco L #ICH260 #ICH260C #ICH260L #ICH260eco #ICH260eco L #ICH750 #ICH750C #ICH750L

Volume: #108 litres #108 litres #256 litres #749 litres

Climate chamber for stability testing in accordance with ICH – WHO – EMA – ASEAN – GMP – GLP – GCCP

Climate chamber ICHeco for stability testing in accordance with ICH and GMP

Climate-friendly and powerful: the ICHeco climate chamber for the stability testing of pharmaceuticals, cosmetics or food works with environmentally-friendly CO2 (R744) as refrigerant. Memmert is thus initiating the climate-friendly future of compressor-cooled temperature control appliances. Furthermore, the ICHeco scores with its unique uniformity of temperature and humidity. Via an air jacket around the hermetically sealed work area, heat is quickly and extensively introduced into the climate chamber over four sides. Icing over of the cooling appliance and sample drying are successfully avoided.

  • digitally controlled active humidification and dehumidification from 10 to 80 % rh
  • Temperature range from – 10 °C to + 60 °C (models ICHeco / ICH), 0 °C to + 60 °C (models ICHeco L / ICH L), + 10 °C to + 50 °C (models ICH C)
  • 1 model variant – TwinDISPLAY
  • double doors as standard: Prevention of contamination and drops or rises in temperature with an optimal view of the chamber load through wide-area interior glass doors

ICH stability testing (Q1B, option 2) in the climate chamber ICHeco L

In the ICHeco L climate chamber, stability tests in accordance with ICH guidelines Q1B, option 2, can be performed, thanks to an illumination unit. Fluorescent lights with cold white light (daylight: light colour 865, 6,500 K) and UV radiation in the spectral range of 320 – 400 nm. Daylight and UV light comply with standard illuminant D65.

Controlled CO2 input in the climate chamber ICH

The climate chamber ICH C offers ideal conditions, especially to test building materials and concrete. Thanks to the electronic CO2 control with automatic neutral position (NDIR measuring system) and an adjustment range of 0 -10 %, you can use it to carry out long-term tests to determine the depth of carbonation in concrete without bulky, costly and power-guzzling test chambers.

Climate-friendly refrigerant CO2 ensures better cooling performance

The climate chamber ICHeco is extremely powerful and environmentally-friendly. Compared to appliances with refrigerant R134a, it scores with faster change speeds during cooling. The Memmert climate chambers ICH / ICH L / ICH C with refrigerant R134a will be available in parallel.

A Memmert ICHeco climate chamber offers unique advantages

  • CO2 Refrigerant (R744) is climate-friendly
  • intuitive and easy-to-use operating menu
  • temperature and humidity ramps programmable via AtmoCONTROL software
  • almost exclusive use of high-quality, corrosion-resistant and easily cleanable stainless steel for the working chamber and housing
  • a wide range of options for programming and documentation using interfaces, integrated data logger and software AtmoCONTROL
  • 3 years guarantee worldwide

Temperature-humidity working range ICHeco

Not all climate chambers are the same. The humidity content of the chamber load, the ambient conditions and the respective temperature-humidity working range are decisive factors in the selection of the right appliance. In the adjacent diagram, you can see the possible temperature/humidity combinations for our climate chamber ICHeco.

Within the respective temperature-humidity range, condensation-free permanent operation is possible. To which extent condensation may occur in the threshold range depends on the humidity content of the chamber load and the ambient conditions.

This seems not to be the appropriate device for your temperature-humidity-working range?
See the other Memmert climate chambers.

#Humidity Chamber HCP #Constant climate chamber HPP #Cleanroom constant climate chamber HPP1060 #Environmental test chambers CTC / TTC #internal dimensions 108 litres, 560 x 480 x 400 mm #internal dimensions 108 litres, 560 x 480 x 400 mm #internal dimensions 108 litres, 560 x 480 x 400 mm #internal dimensions 256 litres, 640 x 800 x 500 mm #internal dimensions 256 litres, 640 x 800 x 500 mm #internal dimensions 256 litres, 640 x 800 x 500 mm #internal dimensions 256 litres, 640 x 800 x 500 mm #internal dimensions 749 litres, 1040 x 1200 x 600 mm #internal dimensions 749 litres, 1040 x 1200 x 600 mm #internal dimensions 749 litres, 1040 x 1200 x 600 mm #internal dimensions 749 litres, 1040 x 1200 x 600 mm #memmert ICH110 Climate Chamber #memmert ICH110C Climate Chamber with CO2 control #memmert ICH110L Climate Chamber with light #memmert ICH110eco Climate Chamber #memmert ICH110eco L Climate Chamber with light  #memmert ICH260 Climate Chamber  #memmert ICH260C Climate Chamber with CO2 control #memmert ICH260L Climate Chamber with light  #memmert ICH260eco Climate Chamber  #memmert ICH260eco L Climate Chamber with light  #memmert ICH750 Climate Chamber #memmert ICH750C Climate Chamber with CO2 control #memmert ICH750L Climate Chamber with light #memmert ICH750eco Climate Chamber #memmert ICH750eco L Climate Chamber with light #ICH110 Climate Chamber #ICH110C Climate Chamber with CO2 control #ICH110L Climate Chamber with light #ICH110eco Climate Chamber #ICH110eco L Climate Chamber with light #ICH260 Climate Chamber #ICH260C Climate Chamber with CO2 control #ICH260L Climate Chamber with light #ICH260eco Climate Chamber #ICH260eco L Climate Chamber with light #ICH750 Climate Chamber #ICH750C Climate Chamber with CO2 control #ICH750L Climate Chamber with light #ICH750eco Climate Chamber #ICH750eco L Climate Chamber with light #108 litres climate chamber #256 litres climate chamber #749 litres climate chamber #ICH110 #ICH110C #ICH110L #ICH110eco #ICH110eco L #ICH260 #ICH260C #ICH260L #ICH260eco #ICH260eco L #ICH750 #ICH750C #ICH750L #108 litres #108 litres #256 litres #749 litres

Humidity chamber, HCP

Humidity chamber, HCP Series, 56L to 241L

$0.00

#Humidity chamber  #Humidity Cabinet

Model:#HCP50 #HCP105 #HCP150 #HCP240

Volume: #56L #107L #156L #241L

Active humidity control creates a controlled environment for accelerated life tests and 85/85 tests.

Perfect environmental simulation in the humidity chamber

This climate chamber offers the full range of comfort, reliability and safety. Thanks to active humidity control from 20 % to 95 % rh and unsurpassed homogeneity of temperature and humidity over the entire interior, it is ideally suited for environmental testing, environmental simulation, accelerated service life tests and 85/85 tests according to IEC 60068-2-67 and IEC 60068-2-78.

  • Temperature range up to +90 °C
  • 4 model sizes (56 to 241 litres volume)
  • 1 model variant: TwinDISPLAY
  • Active humidity control 20 to 95 % rh
  • Double doors as standard: Prevention of contamination and drops or rises in temperature, and at the same time an optimal view of the sensitive chamber load through wide-area heated inner glass door

Humid environment without condensation

The active humidity control ensures an ideal homogeneity of temperature and humidity as well as short recovery times after opening the door in this attractively priced and practically maintenance-free climate chamber. In combination with heating on all six sides, including heated inner glass door, it also minimises vaporisation in the interior and thus the risk of condensed water dripping onto the test object. An aluminium thermal conductive layer supports optimum temperature distribution and serves as a heat accumulator in case of a temporary power failure.

Comfortable equipment for accelerated service life tests

To ensure that environmental tests and accelerated life tests such as 85/85 tests run reliably and safely, the humidity chamber HCP offers a wide range of comfort functions: Standard feed-through on the rear panel, battery-buffered ControlCOCKPIT (option), process runtime starts with SetpointWAIT function only when the setpoint temperature is reached, alarm messages can be sent via e-mail or SMS with MobileALERT (option) and much more.

A Memmert humidity chamber offers unique advantages

  • intuitive and easy-to-use operating menu
  • Heating concept specifically adapted for a precise and homogenous temperature control
  • A wide range of options for programming and documentation of different temperature/humidity profiles via Software AtmoCONTROL
  • Alarm functions, three-fold overtemperature protection and simultaneous display of all parameters provide safety for samples and cell cultures
  • almost exclusive use of high-quality and easily cleanable stainless steel for the working chamber and housing
  • 3 years guarantee worldwide

Temperature Humidity Working Range HCP

Not all climate chambers are the same. The humidity content of the chamber load, the ambient conditions and the respective temperature-humidity working range are decisive factors in the selection of the right appliance. In the adjacent diagram, you can see the possible temperature/humidity combinations for our humidity chambers HCP.

Within the respective temperature-humidity range, permanent operation is possible (at an ambient temperature of 22 °C ± 3 K, relative humidity < 50 %). Condensation may occur in the threshold range. To which extent depends on the humidity content of the chamber load and the ambient conditions.

It goes without saying that every climate chamber has a climate defined by temperature and humidity. However, Memmert climate chambers provide extra reliability during stability testing, climate testing or conditioning and are perfectly designed for the simulation of environmental conditions. In compliance with standards. Reproducible. Economical.

In each Memmert climate chamber, temperature and humidity are distributed homogenously and stably over the entire test period. Each Memmert climate chamber meets the strict requirements of DIN 12880:2007-05 and has a maximum number of safety functions as well as various possibilities for programming and documentation.

#Climate chamber #Humidity chamber #Climate chambers #Humidity chambers #HCP #Humidity chamber #memmert Climate chamber #memmert Humidity chamber #Climate chamber & humidity chamber #Climate chambers & humidity chambers #Climate chamber and humidity chamber #Climate chambers and humidity chambers

#HCP50 Humidity chamber #HCP105 Humidity chamber #HCP150 Humidity chamber #HCP240 Humidity chamber #HCP50 humidity chamber #HCP105 humidity chamber #HCP150 humidity chamber #HCP240 humidity chamber, HCP Humidity chamber# Cabinet #HCP Cabinet #HCP humidity cabinet #memmert humidity chamber

 

Incubator -32L to 749L

$0.00

#memmert #incubator #32 L #749 L #+20 °C #+80 °C

Model:

#IN30plus #IF30 #IF30plus #IN55 #IN55plus #IF55 #IF55plus #IN75 #IN75plus #IF75 #IF75plus #IN110 #IN110plus #IF110 #IF110plus #IN160 #IN160plus #IF160 #IF160plus #IN260 #IN260plus #IF260 #IF260plus #IN450 #IN450plus #IF450 #IF450plus #IN750 #IN750plus #IF750 #IF750plus #ICO50med IVF #ICO105med IVF #IPS260 #IPS750 #ICP110 #ICP110eco #ICP260 #ICP260eco #ICP450 #ICP450eco #ICP750 #ICP750eco #IPP30 #IPP30plus #IPP55 #IPP55plus #IPP110 #IPP110plus #IPP260 #IPP260plus #IPP410 #IPP410plus #IPP750 #IPP750plus #IPP1060 #IPP1060plus

Volume:-

#32 litres #53 litres #74 litres #108 litres #161 litres #256 litres #384 litres #449 litres #749 litres #56 litres #107 litres #1060 litres

The incubator is at home everywhere in the world of research, medicine, pharmaceutics and food analytics, as well as food chemistry.

Highly sensitive, extremely accurate: Memmert lab incubator I

In the laboratory, the Memmert incubator I is ideal for all applications with temperatures up to +80°C – specifically for incubating living cultures at +37 °C. Thanks to the finely-tuned control technology, critical temperature overshoots are completely ruled out, valuable loads are therefore warmed up particularly carefully in this highly precise microbiological incubator.

  • temperature range up to +80 °C
  • 8 model sizes (32 to 749 litres volume)
  • 2 model variants: SingleDISPLAY and TwinDISPLAY
  • natural convection or forced air circulation N/F
  • double doors (interior glass, exterior stainless steel) for a clear view without a drop in temperature
  • activation of sterilisation routine via ControlCOCKPIT(TwinDISPLAY)

Memmert incubator – a gentle force in the laboratory

To guarantee optimum air circulation in the incubator chamber even when it is fully loaded, both model variants SingleDISPLAY and TwinDISPLAY are available with an adjustable air fan. Since our unique all-round surface heating is optimally geared to both natural convection and forced air circulation, the fan can be switched off entirely for sensitive samples and for samples which may not be exposed to air turbulence.

A Memmert incubator offers unique advantages

  • intuitive and easy-to-use operating menu
  • air exchange rates and air flap position electronically controllable
  • temperature ramps programmable via AtmoCONTROL software (models with TwinDISPLAY)
  • pre-heated fresh air prevents temperature fluctuations
  • almost exclusive use of high-quality, corrosion-resistant and easily cleanable stainless steel for the working chamber and housing
  • heating concept specifically adapted for a precise and homogenous temperature control
  • a wide range of options for programming and documentation using interfaces, integrated data logger and software AtmoCONTROL

3 years guarantee worldwide

Incubator, IN55, 53L TFT digital display, 230 V, 50/60 Hz, Germany

$0.00

The heating of this incubator is optimally tuned for natural convection and valuable chamber loads for research,
pharmaceutics, medicine and food chemistry are warmed up very carefully. On this page, you can find all the
essential technical data on our incubator. Our customer relations team will be pleased to help if you want further
information.

Incubator with natural air circulation for the daily heating requirements

With the unique TFT colour display with touch-sensitive buttons.

  • Temperature from 5°C above ambient up to 80°C
  • Internal chamber in stainless steel
  • Fully insulated stainless steel door, with internal glassdoor
  • Adjustment of pre-heated fresh air admixture by air flap control in 10% steps
  • Multifunctional digital PID microprocessor controller
  • Parameters adjustable on ControlCOCKPIT: Temperature, air flap position and programme time
  • Resolution of display for setpoint and actual value: 0.1°C
  • Adjustable digital timer: from 1 min. to 99 days, 23 hours
  • SetpointWAIT function – time starts when the desired temperature is reached
  • Fast sterilisationprogram (4 hours/160°C) for sterilisation of the chamber
  • Internal data logger with capacity for at least 10 years.
  • Ethernet interface for sending of the protocol log, to for example. PC software AtmoCONTROL
  • Adjustable over-temperature protection, protection class 1 according to DIN 12 880

Specifications

Temperature

Setting temperature range +20 to +80 °C
Working temperature range at least 5 above ambient temperature to +80 °C
Setting accuracy temperature 0.1 °C
Temperature sensor 1 Pt100 sensor DIN class A in 4-wire-circuit

Control technology

Language setting German, English, Spanish, French, Polish, Czech, Hungarian
ControlCOCKPIT SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable temperature values
adjustable parameters temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime

Ventilation

Convection natural convection
Fresh air Admixture of pre-heated fresh air by electronically adjustable air flap
Vent vent connection with restrictor flap

Communication

Documentation programme stored in case of power failure
Programming AtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand).

Safety

Temperature control adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
Autodiagnostic system for fault analysis

Standard equipment

Door fully insulated stainless steel door with 2-point locking (compression door lock)
Internals 1 stainless steel grid(s), electropolished
Works calibration certificate incl. works calibration certificate for +37°C
Door inner glass door

Stainless steel interior

Interior easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Volume 53 l
Dimensions w(A) x h(B) x d(C): 400 x 400 x 330 mm
Max. number of internals 4
Max. loading of chamber 80 kg
Max. loading per internal 20 kg

Textured stainless steel casing

Dimensions w(D) x h(E) x d(F): 585 x 784 x 514 mm (d +56mm door handle)
Housing rear zinc-plated steel

Electrical data

Voltage Electrical load 230 V, 50/60 Hz
approx. 1000 W
Voltage Electrical load 115 V, 50/60 Hz
approx. 900 W

Ambient conditions

Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Altitude of installation max. 2,000 m above sea level
Ambient temperature +5 °C to +40 °C
Humidity rh max. 80 %, non-condensing
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 730 x 950 x 670 mm
Net weight approx. 57 kg
Gross weight carton approx. 76 kg
Categories:  

Climate chamber ICH

$0.00

Its excellent temperature and humidity homogeneity positions the climate chamber ICH in the international top league of stability test chambers. Thermal transfer to the interior – whether heat or cold – is guaranteed through a large-area air jacket on four sides. Thanks to its hermetically sealed interior, the climate chamber ICH does not consume any water once the setpoints have been reached, provided that the door is not opened. Two high-grade platinum temperature sensors with mutual monitoring guarantee interruption-free long-term operation even if one sensor should fail.

  • 3 models
    • Climate chamber ICH 110/260/750 with temperature and humidity
    • Temperature range from – 10 °C to + 60 °C  – ICH 110L/260L/750L with light
    • Temperature range from 0 °C to + 60 °C – ICH 110C/260C/750C with CO2 control
    • Temperature range from 0 °C to + 60 °C – 3 model sizes (108 to 749 litres volume)
  • 1 model variant
  • TwinDISPLAY
  • Digitally controlled active humidification and dehumidification from 10 to 80 % rh
  • Double doors as standard
    • Prevention of contamination and drops or rises in temperature with an optimal view of the chamber load through wide-area interior glass doors

In the ICH L model, tests in accordance with ICH guidelines Q1B, option 2, can be performed, thanks to an illumination unit. Fluorescent lights with cold white light (daylight: light colour 865, 6,500 K) and UV radiation in the spectral range of 320 – 400 nm. Daylight and UV light comply with standard illuminant D65. Thanks to a combination of temperature, humidity and digital CO2 control, the climate chamber ICH C is optimally suited for building materials testing or applications in cell biology requiring temperatures below room temperature. The air jacket temperature control ensures that the working chamber is sealed off from the outside all round.

The advantages:

  • Rapid and precise temperature control
  • No evaporator-dependent drying out of samples
  • No icing over of the cooling unit
  • No dehumidification of the working chamber

With the help of our model selection, with dimensioned model sketches and extensive technical data for download, you will find the perfect climate chamber ICH for your needs. For work predominantly close to the ambient temperature, we recommend the extremely energy-saving Memmert constant climate chamber HPP with an innovative and eco-friendly Peltier heating and cooling system.

A Memmert climate chamber ICH offers unique advantages:

  • Intuitive and easy-to-use operating menu
  • Temperature and humidity ramps programmable via AtmoCONTROL software
  • Almost exclusive use of high-quality, corrosion-resistant and easily cleanable stainless steel for the working chamber and housing
  • Peltier heating and cooling concept specifically adapted for a precise and homogenous temperature control
  • A wide range of options for programming and documentation using interfaces, integrated data logger and software AtmoCONTROL
  • 3 years guarantee worldwide
  • The temperature tests for all climate chambers are performed in accordance with DIN 12880:2007-05
  • Each climate chamber ICH is safety-tested
Categories: