Elite Scientific and Meditech Co.

elitetradebd@gmail.com

+8801711196910

elite, elitetrade, elitetradebd, elitescientific, elitebd, elitescientificbd, elitescientificco, elitescientificandmeditechco, elitescientificmeditechco, elitescientific&meditechco, elite scientific, elite scientific meditech co, elite scientific & meditech co, elite scientific and meditech co, scientific company in bangladesh, be scientific co, Bangladeshi scientific co, scientific products seller in Bangladesh, Scientific products seller in bd, scientific products supplier in bd, laboratory products selller in bd, lab goods seller in bd, Chemicals supplier in bd, Chemicals supplier in Bangladesh, glassware seller in bd, scientific goods price in bd, Tender goods seller in bd

Liquid Particle Counter

$0.00

Liquid Particle Counter LLPC-A10 comes with high precision syringe sampling and semiconductor laser sensor. Designed using photoresist method principle, it detects size and number of solid particles in a liquid sample. It�s a high quality performance instrument, designed in accordance with international standards such as ISO4406:1999.

Features

  • Precision syringe sampling with stable sampling rate
  • Built-in threshold- diameter curve
    8 Measurement channels
  • Wide Detection range
  • Large LCD touch screen display
  • Large data processing, classification, verification
  • Built � in printer RS232 interface with connection to printer
  • Instrument calibration and verification as per international standards

Specifications

Measurement range
1 to 250 µm (25 sensors)
1 µm to 320 µm (32 sensors)
2 µm to 400 µm (40 sensors)
Light Source Semiconductor laser light
Measurement Channels 8 channels
Particle size 1 µm (2 µm) to arbitrarily set within 100 µm
Sensitivity
1 µm (ISO04402)
4 µm (C) (ISO11171) (25, 32 sensor)
2 µm (ISO4402) (40 sensor)
Sample volume ≥ 0.5 ml
Sample accuracy Better than ± 1 %
Sampling rate 5 ml/ min to 60 ml/ min
Resolution < 10 %
Coincidence error limit 10,000/ ml (5 % Coincidence error)
Maximum cabin pressure 0.6 MPa
Maximum cabin vacuum 0.08 MPa
Maximum viscosity detection
350 cps (40 sensor)
250 cps (32 sensor)
100 cps (25 sensor)
Ambient temperature 10 to 40°C
Interface RS232
Dimension 220 × 380 × 480 mm
Weight 28 kg
Power 100 to 245 V, 50 Hz, ± 1 %, , 80 W