#i2 Visible Spectrophotometer #i3 UV-VIS Spectrophotometer #i5 UV-VIS Spectrophotometer#i6 UV-VIS Spectrophotometer #i7 UV-VIS Spectrophotometer #i8 Double-Beam UV-VIS Spectrophotometer #i9 Double-Beam UV-VIS Spectrophotometer
Description
1.Standard scanning software can directly complete functions of Quantitative;Kinetics;Wavelength Scan; Multi Wavelength; DNA/Protein and Data processing.
Specification
i2 | i3 | i5 | i6 | i7 | i8 | i9 | |
Wavelength Range | 320-1100nm | 190-1100nm | 190-1100nm | 190-1100nm | 190-1100nm | 190-1100nm | 190-1100nm |
Bandwidth | 2nm | 2nm | 1.8nm | 1.8nm | 0.5/1/2/4nm | 1.8nm | 0.5/1/2/4/5nm |
Wavelength Accuracy | ±0.5nm | ±1nm | ±0.5nm | ±0.1nm (D2 656.1nm); ±0.3nm (Full range) | |||
Wavelength Reproducibility | ≤0.2nm | ≤0.3nm | ≤0.2nm | ≤0.1nm | ≤0.1nm | ≤0.1nm | ≤0.1nm |
Photometric Accuracy | ±0.3%T | ±0.5%T | ±0.3%T | ±0.2%T | ±0.2%T | ±0.2%T | ±0.2%T |
Photometric Repeatability | 0.15% T | ≤0.2%T | 0.15% T | 0.15% T | 0.15% T | 0.15% T | 0.15% T |
Straylight | ≤0.05%T | ≤0.05%T | ≤0.05%T | ≤0.03%T | ≤0.03%T | ≤0.03%T | ≤0.03%T |
Stability | ±0.0001A/h(500nm) | ±0.001A/h(at 500nm) | ±0.0004A/h(at 500nm) | ||||
Baseline Flatness | ±0.001A/h | ±0.0005A | ±0.001A | ±0.0015A | ±0.0015A | ±0.001A | ±0.001A |
Noise | ±0.0005A/h | ±0.001A | ±0.0005A | ±0.0005A | ±0.0005A | ±0.0005A | ±0.0005A |
Photometric Range | “0-200%T, -0.3–3A, 0-9999C | 0-200%T, -4.0–4.0A, 0-9999C | |||||
Wavelength setting mode | Automatic | ||||||
Scanning speed | — | — | High,Middle,Low,Optional | ||||
Output | USB Port | ||||||
Printer port | Parallel Port | ||||||
Display | LCD(320*240) | ||||||
Light Source | Tungsten Lamp | Deuterium&Tungsten Halogen Lamp | |||||
Detector | Silicon Photodiode | ||||||
Power | 220V AC ±10%/50Hz or 110V AC / 60Hz | ||||||
Dimension | 460x380x180mm | 420×300 x160mm | 460×380 x180mm | 470×370 x180mm | 470×370 x180mm | 625×430 x210mm | 625×430 x210mm |
Weight | 15Kg | 13Kg | 20Kg | 20Kg | 20kg | 28kg | 28Kg |
#T860 automatic titrator
Description
Hanon T860 Automatic Titrator is a lab titration device with high analysis accuracy under the principle of potentiometric titration. It enjoys modular design, consist of three parts: volume titration device, control device and measuring device. The titration methods include macro titration, micro titration, volume setting titration, EP, etc.
Characteristics
Zero-dead volume solenoid valve,changing easily for PTFE burette.
Ultrathin stirring device, adopt coil realize magnetic Stir.
Accuracy closed-loop control for volume.
High-accuracy burette accurate to 0.005mm.
Simple design, discrete stirring unit, detachable easily.
LCD touch screen.
Wide operating voltage range 110~240V, apply to global client.
Specification
Measurement mode | MEAS | pH/mV |
CAL | pH calibration | |
SET | End point setting | |
MET | monotonic titration | |
Burette | Volume | 15mL/25mL |
Stirring | Mode | magnetic |
Titration cup volume | 100mL | |
Electrode | mv/pH | 1 x BNC |
Reference | 1 | |
PT1000 | 1 x NTC | |
Communication | USB | 1 |
Stirrer | 1 | |
Power | 1 | |
Printer | 1 |
Technical
mV/pH | Measurement range | ±1,999 mV 0.00~14.00 pH |
Resolution: | 0.1 mV / 0.01 pH | |
Error | 0.2 mV / 0.02pH | |
PT1000 | Measurement range | 0-100℃ |
Resolution: | 0.1 ℃ | |
Error | 0.2 ℃ | |
Burette | 15 mL tolerance | ±0.025mL |
25 mL tolerance | ±0.035mL | |
Dosing Repeatability | 0.20% | |
mV Repeatability | ≤0.2mV | |
Controller | Display | 7” inch touch screen |
Information | Weight | 10 Kg |
Power | 110-240V,50/60 Hz |
#T960 automatic titrator
Characteristics:
Description
Hanon T960 Automatic Titrator meet the demand of mult complex liquid inlet automatically. There is no need to add solution manually. With the most secure built-in burette, to avoid contact with harmful chemicals which protect user security; Cloud services , connect via the Ethernet port ,to achieve the sharing between the instrument and Hanon huge database.Lead the laboratory instruments to Cloud generation.
Features:
Mixed synchronous inlet system:
The world’s leading Mixed synchronous inlet system with 4-channel sample inlet simultaneously, supports expand up to 6 channels. Totally automatic in add titrant, any combination of additives or auxiliary agents.
Built-in burette:
The most secure built-in burette, burette built in the instrument and can be observed though the window during the test to avoid the harmless chemical leak. Burette volume is optional:5ml/15ml/25ml
Cloud Service:
With ethernet interface,it can storage the database to the cloud server.Lead the laboratory instruments to Cloud generation.
Titration platform diversification:
Flexible to work with independence titration stand or 16 auto-sampler, multi-tasking operating through 16 auto-sampler can be fully automated, the whole process without staff to monitor, improve the work much more efficiency.
Specification
T960 BASIC | T960 PRO | T960 PLUS | |
MV measuring range | ±2000MV | ● | ● |
MV measuring Resolution | Resolution 0.1MV | ● | ● |
Accuracy 0.1MV | |||
pH measuring range | -20.000PH~+20.000PH | ● | ● |
pH measurement | Resolution 0.001PH | ● | ● |
Accuracy 0.003PH | |||
Temperature measuring range | -5~120℃ | ● | ● |
Temperature measuring accuracy | ±0.1℃ | ● | ● |
Interface | 2xUSB, RS232, Ethernet | ● | ● |
Max titration station | 2 | 4 | 6 |
Max liquid filling module | 2 | 4 | 4 |
Burette | 5ml, 10ml, 25ml | ● | ● |
Burette resolution | 1/48,000 | ● | ● |
Time the burette charging | 16 seconds (100% charging speed) | ● | ● |
PC control | ● | ● | ● |
Electrodeconnector type | mv/pHmeasurement electrode connetor、reference electrode connector 、PT1000 temperature electrode connector | ● | ● |
Auto-Sampler | 16 positions (Optional) | ○ | ○ |
#MP490 melting point apparatus
MP420/MP430 Video Melting Point Apparatus, and MP450/MP470/MP490 Automatic Video Melting Point Apparatus, perfectly integrate video technology into melting point measuring. It not only provides users with reliable melting point test result, but also intuitively displays temperature curve and real-time video image, and achieves the function of save, replay the video,stamp video and so on. It also provides better help for users to observe melting process of sample.
Features
Visual and Automatic Determination
It includes a color display to observe the phase transition, is equipped with a video camera and a video replay function which allows reviewing the melting point determination. It also suit for color samples, which tradition automatic melting point apparatus not available.
Available Testing 4 Samples Concurrently
Max concurrent sessions 4 samples each batch, record and track each result independently, meet FDA 21CFR part 11 requirements.
Multi-user ID Classify Function
Adopt User ID management function. Users ID are available for adding, deleting and jurisdiction setting.
All Screen Design
Adopt a 8 inch capacitive touch screen, with the brilliant Android system, let the operating easier.
Amazing Data Processing & Exporting
The methods, atlas and detailed result are available for storage. Connecting printer or all data export are both available.
Compliant with GLP, GMP(EU,UPS&CN) and FDA 21CFR part 11
Specification
MP420 | MP430 | MP450 | MP470 | MP490 | |
Position Capacity | 4 | 4 | 4 | 4 | 4 |
Magnification of lens | 8 | 8 | 8 | 8 | 8 |
Display | 8 inch capacitive touch screen | 8 inch capacitive touch screen | 8 inch capacitive touch screen | 8 inch capacitive touch screen | 8 inch capacitive touch screen |
Temp Range | 300℃ | 350℃ | 350℃ | 400℃ | 400℃ |
Temperature resolution | 0.1℃ | 0.1℃ | 0.1℃ | 0.1℃ | 0.1℃ |
Temp Accuracy | ±0.4℃(<200℃)±0.7℃(<300℃) | ±0.4℃(<200℃)±0.7℃(<300℃) | ±0.4℃(<200℃)±0.7℃(<300℃) | ±0.4℃(<200℃)±0.7℃(<300℃) | ±0.4℃(<200℃)±0.7℃(<300℃) |
Temp Gradients | Step-less setting at 0.1℃–20℃ | Step-less setting at 0.1℃–20℃ | Step-less setting at 0.1℃–20℃ | Step-less setting at 0.1℃–20℃ | Step-less setting at 0.1℃–20℃ |
ID classify | No | 20 | 20 | 30 | 100 |
Camera&Video | Camera | Camera&Video | Camera&Video | Camera&Video | Camera&Video |
Video Replay | Manual | Manual | Automatic | Automatic | Automatic |
Model | No | Yes | Yes | Yes | Yes |
Graphic | 0 | 0 | 100 | 200 | 500 |
Methods Storage Capacity | 0 | 30 | 30 | 80 | 200 |
Rom Capacity | 8GB | 16GB | 16GB | 32GB | 128GB |
FDA CFR21 part 11 | No | Yes | No | Yes | Yes |
Communication Ports | USB×3,RS232,Wi-Fi,RJ45 | USB×3,RS232,Wi-Fi,RJ45 | USB×3,RS232,Wi-Fi,RJ45 | USB×3,RS232,Wi-Fi,RJ45 | USB×3,RS232,Wi-Fi,RJ45 |
#P850 Pro Auto Digital Polarimeter
Features
Peltier Temperature Control
The Peltier system automatically controls the sample temperature in the max range from 10 °C to 50 °C.
User Calibration System
Additional Standard polarization tubes with software for calibration by user.
Large Color Touch Screen
Large color touch screen display and innovative software interface give extremely convenient device operation and data acquisition. P810/P850 adopt a 5.6inch TFT display, P810 Pro/ P850 Pro adopt 8inch TFT display.
LED Light Source
High brightness LED lights with service life exceeding 100000 hours. To void the light source exchanging.
Description
P Series Polarimeter with automatic photoelectric inspection technology, boasts accurate and reliable measurement, convenient operation. By detecting the optical rotation, the density, content and purity of substances can be analyzed and determined. It’ s widely used in pharmaceutical, petroleum, food, chemicals, flavors, spices, sugar marking and other industries , and universities and research institutes.
New upgraded models: P810 Pro/ P850 Pro adopt brilliant Android system and communication ports, let new models meet FDA CFR21 part 11.
New P810 Pro/ P850 Pro New Characteristics
Adopt a 8 inch TFT touch screen, with the brilliant Android system, let the operating easier.
Meet FDA CFR21 part11.
User ID management function. Users ID are available for adding, deleting and jurisdiction setting.
Timing measuring and atlas display function, displaying the samples optical rotation changing dynamic process.
8GB room for data storage.
Classify and synchronous result searching function is convenience for user’s result searching.
3 USD port for U-Disk, keyboard & mouse connecting. And WiFi for connecting printer. Let operating much easier.
Specification
P810 | P850 | P810 Pro | P850 Pro | |
Measuring range | ±89.99°Arc; ±259°Z | ±89.99°Arc; ±259°Z | ±89.99°Arc; ±259°Z | ±89.99°Arc; ±259°Z |
Measuring Model | Optical Rotation; Specific Rotation; Concentration; Sugar Degree | Optical Rotation; Specific Rotation; Concentration; Sugar Degree | Optical Rotation; Specific Rotation; Concentration; Sugar Degree | Optical Rotation; Specific Rotation; Concentration; Sugar Degree |
Min.reading | 0.001°(optical rotation) | 0.001°(optical rotation) | 0.001°(optical rotation) | 0.001°(optical rotation) |
Accuracy | ±0.01º(-45º. optical rotation ≤+45º; ±0.02º ( optical rotation<﹣45º or optical rotation>﹢45º) | ±0.01º(-45º. optical rotation ≤+45º; ±0.02º ( optical rotation<﹣45º or optical rotation>﹢45º) | ±0.01º(-45º. optical rotation ≤+45º; ±0.02º ( optical rotation<﹣45º or optical rotation>﹢45º) | ±0.01º(-45º. optical rotation ≤+45º; ±0.02º ( optical rotation<﹣45º or optical rotation>﹢45º) |
Repeatability (standard deviation) | 0.002°(optical rotation) | 0.002°(optical rotation) | 0.002°(optical rotation) | 0.002°(optical rotation) |
Min. Transmittance | 1% | 1% | 1% | 1% |
Light Source | LED | LED | LED | LED |
Wavelength | 589.3nm LED | 589.3nm LED | 589.3nm LED | 589.3nm LED |
Temperature Control | No | Peltier | No | Peltier |
Temperature Control Range | 15℃~30℃ | 10℃~50℃ | ||
Temperature Control Accuracy | 0.3℃ | 0.3℃ | ||
Communication Port | USB; RS232 | USB; RS232 | USB×3,RS232,Wi-Fi | USB×3,RS232,Wi-Fi |
Data Storable | 1000 | 1000 | 8GB | 8GB |
21CFR Part 11 | No | No | Yes | Yes |
Display | 5.6inch | 5.6inch | 8inch | 8inch |
Sample Cell | Upto 200mm | Upto 200mm | Upto 200mm | Upto 200mm |
Graphic | No | No | 4hours | 4hours |
Power | 220V AC (±10%) 50Hz | 220V AC (±10%) 50Hz | 220V AC (±10%) 50Hz | 220V AC (±10%) 50Hz |
Calibration System | Option | Option | Option | Option |
#P850 Auto Digital Polarimeter
Features
Peltier Temperature Control
The Peltier system automatically controls the sample temperature in the max range from 10 °C to 50 °C.
User Calibration System
Additional Standard polarization tubes with software for calibration by user.
Large Color Touch Screen
Large color touch screen display and innovative software interface give extremely convenient device operation and data acquisition. P810/P850 adopt a 5.6inch TFT display, P810 Pro/ P850 Pro adopt 8inch TFT display.
LED Light Source
High brightness LED lights with service life exceeding 100000 hours. To void the light source exchanging.
Description
P Series Polarimeter with automatic photoelectric inspection technology boasts accurate and reliable measurement, convenient operation. By detecting the optical rotation, the density, content and purity of substances can be analyzed and determined. It’ s widely used in pharmaceutical, petroleum, food, chemicals, flavors, spices, sugar marking and other industries , and universities and research institutes.
New upgraded models: P810 Pro/ P850 Pro adopt brilliant Android system and communication ports, let new models meet FDA CFR21 part 11.
New P810 Pro/ P850 Pro New Characteristics
·Adopt a 8 inch TFT touch screen, with the brilliant Android system, let the operating easier.
·Meet FDA CFR21 part11.
·User ID management function. Users ID are available for adding, deleting and jurisdiction setting.
·Timing measuring and atlas display function, displaying the samples optical rotation changing dynamic process.
·8GB room for data storage.
·Classify and synchronous result searching function is convenience for user’s result searching.
·3 USD port for U-Disk, keyboard & mouse connecting. And WiFi for connecting printer. Let operating much easier.
Description
P Series Polarimeter with automatic photoelectric inspection technology, boasts accurate and reliable measurement, convenient operation. By detecting the optical rotation, the density, content and purity of substances can be analyzed and determined. It’ s widely used in pharmaceutical, petroleum, food, chemicals, flavors, spices, sugar marking and other industries , and universities and research institutes.
New upgraded models: P810 Pro/ P850 Pro adopt brilliant Android system and communication ports, let new models meet FDA CFR21 part 11.
New P810 Pro/ P850 Pro New Characteristics
Adopt a 8 inch TFT touch screen, with the brilliant Android system, let the operating easier.
Meet FDA CFR21 part11.
User ID management function. Users ID are available for adding, deleting and jurisdiction setting.
Timing measuring and atlas display function, displaying the samples optical rotation changing dynamic process.
8GB room for data storage.
Classify and synchronous result searching function is convenience for user’s result searching.
3 USD port for U-Disk, keyboard & mouse connecting. And WiFi for connecting printer. Let operating much easier.
Specification
P810 | P850 | P810 Pro | P850 Pro | |
Measuring range | ±89.99°Arc; ±259°Z | ±89.99°Arc; ±259°Z | ±89.99°Arc; ±259°Z | ±89.99°Arc; ±259°Z |
Measuring Model | Optical Rotation; Specific Rotation; Concentration; Sugar Degree | Optical Rotation; Specific Rotation; Concentration; Sugar Degree | Optical Rotation; Specific Rotation; Concentration; Sugar Degree | Optical Rotation; Specific Rotation; Concentration; Sugar Degree |
Min.reading | 0.001°(optical rotation) | 0.001°(optical rotation) | 0.001°(optical rotation) | 0.001°(optical rotation) |
Accuracy | ±0.01º(-45º. optical rotation ≤+45º; ±0.02º ( optical rotation<﹣45º or optical rotation>﹢45º) | ±0.01º(-45º. optical rotation ≤+45º; ±0.02º ( optical rotation<﹣45º or optical rotation>﹢45º) | ±0.01º(-45º. optical rotation ≤+45º; ±0.02º ( optical rotation<﹣45º or optical rotation>﹢45º) | ±0.01º(-45º. optical rotation ≤+45º; ±0.02º ( optical rotation<﹣45º or optical rotation>﹢45º) |
Repeatability (standard deviation) | 0.002°(optical rotation) | 0.002°(optical rotation) | 0.002°(optical rotation) | 0.002°(optical rotation) |
Min. Transmittance | 1% | 1% | 1% | 1% |
Light Source | LED | LED | LED | LED |
Wavelength | 589.3nm LED | 589.3nm LED | 589.3nm LED | 589.3nm LED |
Temperature Control | No | Peltier | No | Peltier |
Temperature Control Range | 15℃~30℃ | 10℃~50℃ | ||
Temperature Control Accuracy | 0.3℃ | 0.3℃ | ||
Communication Port | USB; RS232 | USB; RS232 | USB×3,RS232,Wi-Fi | USB×3,RS232,Wi-Fi |
Data Storable | 1000 | 1000 | 8GB | 8GB |
21CFR Part 11 | No | No | Yes | Yes |
Display | 5.6inch | 5.6inch | 8inch | 8inch |
Sample Cell | Upto 200mm | Upto 200mm | Upto 200mm | Upto 200mm |
Graphic | No | No | 4hours | 4hours |
Power | 220V AC (±10%) 50Hz | 220V AC (±10%) 50Hz | 220V AC (±10%) 50Hz | 220V AC (±10%) 50Hz |
Calibration System | Option | Option | Option | Option |
Features
Peltier Temperature Control
The Peltier system automatically controls the sample temperature in the max range from 10 °C to 50 °C.
User Calibration System
Additional Standard polarization tubes with software for calibration by user.
Large Color Touch Screen
Large color touch screen display and innovative software interface give extremely convenient device operation and data acquisition. P810/P850 adopt a 5.6inch TFT display, P810 Pro/ P850 Pro adopt 8inch TFT display
LED Light Source
High brightness LED lights with service life exceeding 100000 hours. To void the light source exchanging.
Description
P Series Polarimeter with automatic photoelectric inspection technology, boasts accurate and reliable measurement, convenient operation. By detecting the optical rotation, the density, content and purity of substances can be analyzed and determined. It’ s widely used in pharmaceutical, petroleum, food, chemicals, flavors, spices, sugar marking and other industries , and universities and research institutes.
New upgraded models: P810 Pro/ P850 Pro adopt brilliant Android system and communication ports, let new models meet FDA CFR21 part 11.
New P810 Pro/ P850 Pro New Characteristics
Adopt a 8 inch TFT touch screen, with the brilliant Android system, let the operating easier.
Meet FDA CFR21 part11.
User ID management function. Users ID are available for adding, deleting and jurisdiction setting.
Timing measuring and atlas display function, displaying the samples optical rotation changing dynamic process.
8GB room for data storage.
Classify and synchronous result searching function is convenience for user’s result searching.
3 USD port for U-Disk, keyboard & mouse connecting. And WiFi for connecting printer. Let operating much easier.
Specification
Model |
P810 |
P850 | P810 Pro |
P850 Pro |
Measuring range | ±89.99°Arc; ±259°Z | ±89.99°Arc; ±259°Z | ±89.99°Arc; ±259°Z | ±89.99°Arc; ±259°Z |
Measuring Model | Optical Rotation; Specific Rotation; Concentration; Sugar Degree | Optical Rotation; Specific Rotation; Concentration; Sugar Degree | Optical Rotation; Specific Rotation; Concentration; Sugar Degree | Optical Rotation; Specific Rotation; Concentration; Sugar Degree |
Min.reading | 0.001°(optical rotation) | 0.001°(optical rotation) | 0.001°(optical rotation) | 0.001°(optical rotation) |
Accuracy | ±0.01º(-45º. optical rotation ≤+45º; ±0.02º ( optical rotation<﹣45º or optical rotation>﹢45º) | ±0.01º(-45º. optical rotation ≤+45º; ±0.02º ( optical rotation<﹣45º or optical rotation>﹢45º) | ±0.01º(-45º. optical rotation ≤+45º; ±0.02º ( optical rotation<﹣45º or optical rotation>﹢45º) | ±0.01º(-45º. optical rotation ≤+45º; ±0.02º ( optical rotation<﹣45º or optical rotation>﹢45º) |
Repeatability (standard deviation) | 0.002°(optical rotation) | 0.002°(optical rotation) | 0.002°(optical rotation) | 0.002°(optical rotation) |
Min. Transmittance | 1% | 1% | 1% | 1% |
Light Source | LED | LED | LED | LED |
Wavelength | 589.3nm LED | 589.3nm LED | 589.3nm LED | 589.3nm LED |
Temperature Control | No | Peltier | No | Peltier |
Temperature Control Range | 15℃~30℃ | 10℃~50℃ | ||
Temperature Control Accuracy | 0.3℃ | 0.3℃ | ||
Communication Port | USB; RS232 | USB; RS232 | USB×3,RS232,Wi-Fi | USB×3,RS232,Wi-Fi |
Data Storable | 1000 | 1000 | 8GB | 8GB |
21CFR Part 11 | No | No | Yes | Yes |
Display | 5.6inch | 5.6inch | 8inch | 8inch |
Sample Cell | Upto 200mm | Upto 200mm | Upto 200mm | Upto 200mm |
Graphic | No | No | 4hours | 4hours |
Power | 220V AC (±10%) 50Hz | 220V AC (±10%) 50Hz | 220V AC (±10%) 50Hz | 220V AC (±10%) 50Hz |
Calibration System | Option | Option | Option | Option |
#A630 Refractometer
Features:
Automatic Refractometer is widely used on kinds of industries, it is one of the most common instruments in the oil industry, pharmaceutical industry, paint industry, food industry, daily chemical industry, sugar industry, geological exploration or institutions.
Measuring prism is made of high hardness sapphire glass, of excellent properties against corrosion and scratches, free for clean and durable
High-resolution CCD detector offers automatic measurement, avoiding personal errors, having higher accuracy; transparent, translucent liquids or ropiness liquids samples are all available;
LED cold light source with the long working life more than 100000 hours;
Description
Hanon A670/A630 Automatic Refractometer includes high-performance CCD light-sensitive part, it can automatically measure the refractive index(nD) of transparent, translucent, dark, viscous and other types liquid sample, also can measure sugar degree(Brix) of sugar solution. It’s accurate, reliable, fast, easy operation. Full compliance with FDA 21CFR part.11. A670 innovative use of cloud service systems, all the data from instrument can update to cloud server, especially for group company or government customer, automatic refractometer has a wid range of use, within the field of factories, university and related scientific research institute, petroleum, oil, pharmaceutical, paints, food, daily chemical, sugar and geological prospecting industries.
Characteristic
Specification
Refractive index measuring range (nD) | 1.30000-1.70000 |
Measured value error (nD) | ±0.0002 |
Measurement resolution (nD) | 0.0001 |
Scope of sugar solution mass fraction (Brix) | 0-100% |
Measured value error (Brix) | ±0.1% |
Measurement resolution (Brix) | 0.1% |
Temperature control mode | Peltier |
Scope of temperature measuring | 0-100℃ |
Temperature control range | 10-50 ℃ |
Temperature control accuracy | ±0.1℃ |
Interface | RS232/USB/SD card/U disk/Internet |
Data storage capacity | 4GB |
Display modes | 7 inches, TFT touch screen |
Power supply(Power adapter) | 110V~240V(12V,10A) |
Power | 45W |
Net weight | 12kg |
Dimensions | 365mmX300mmX150mm |
Visit us | www.elitetradebd.com |
Supplier | Elite Scientific & Meditech Co |
#HN200 Sample concentrator
Description
Each blowing needle can be controlled independently; the height of gas cavity is adjustable
Characteristics
Sample visualization.
Technical data:
Temperature range | Room temperature +5℃~180℃ |
Heating time | ≤30min (from 40℃ to 180℃) |
Temperature accuracy | ±0.5℃(40-100℃) |
±1℃(100~180℃) | |
Display accuracy | 0.1 |
Display mode | LED display |
Nitrogen flow | 0-10L/min |
Nitrogen pressure | ≤0.1MPa |
Dimensions | 280mmX240mmX500mm |
Supplier | Elite Scientific & Meditech Co |
Visit us | www.elitetradebd.com |
#Automatic solid phase extraction
Sample Pre-treament Series
SPE100/200/400 Automatic Solid Phase Extraction
Characteristic:
The injection needle and the extraction column are sealed by a universal plug sealing structure and equipped with H-axis pressing block to ensure good sealing.
The suspension structure is adopted in the test tube bracket, which ensures the positioning accuracy of the manipulator to the greatest extent.
The instrument standard router can be connected by WiFi or network line, which fully realizes the whole process of automation and unattended.
Using modular design, only one communication line and pipeline can work between modules, and it is convenient to replace and maintain.
Add emergency stop button, when the instrument is abnormal, manual operation can be emergency stop, to maximize user safety.
Handwheel adjusting origin greatly saves the time of adjusting origin.
Main features
SPE100/SPE200/SPE400/SPEX00(Customization)
Channels | 1/2/4/X |
Channel Independent Control | Support |
Sample pump | Precision Continuous Sample Pump |
Pump flow rate | 0.1-120mL/min |
Types of solvents | 8Seed, 500 mL Solvent Bottle |
Sample volume | 0.1mL-20L |
Sample tube volume | 10 – 20 mL,customization |
Receiving volume | 10mL/15mL, other specifications customized |
Puncture sampling | Supporting |
Loading mode | X/Y/Z Three-Dimensional Manipulator |
Accuracy of manipulator | 0.01mm |
Sealing technology | Docking Technology of Independent Seal Cover and Pan-plug Seal |
Extraction mode | Positive pressure extraction |
column pressure | ≦80psi |
Pressure sensing | Support, Overvoltage Alarm, Threshold Editable (5-145psi) |
Column drying method | Nitrogen purge |
Waste collection | Branch collection,3 channels |
A/B in series | Support, A/B Pillar |
Cleaning mode | Nitrogen purge, solvent cleaning, blowing stirring cleaning and other modes are |
used to clean the inner and outer walls of sample needles. | |
Liquid level following | Support |
Solvent Residual Capacity | Solvent deficiency hints |
Host Size (H*W*D) | 680mm x 580mm x 670mm |
Voltage | 220/110V,50/60Hz |
working temperature | 5-40Temperature |
Working humidity | 0-90% without condensation |
power | 60W/80W/120W/<200W |
#SOX406 fat analyzer
Feature:
· Integral metal heating, while scope and high precision of temperature control
· Electric circuit is isolated from the extraction space, ensuring devices security
· Over-temperature alarming and timer reminding functions are available
· Timer circuit is isolated from the extraction space, ensuring device security
· Triple alarms i.e. sound, light, LCD screen word prompts are available
· Abundant interface contents give simultaneous display of given temperature, actual temperature, given time and heating time
· The lifting connection of linear bearing conduction technique gives smooth and comfortable lifting operation
· Intelligent man-machine dialogue control system
· Exclusive air insulation technique leaves the case in room temperature, has thermal insulation and temperature maintenance two functions
· 4.3″ LCD screen and microcomputer control system are adopted
Description
SOX406 Fat Analyzer is base on the Soxhlet extraction principle and integrates such functions as soaking, extraction, leaching, heating, condensation and solvent recovery. It features sealed metal bath heating with automatic temperature control, ensuring uniform heating and safe operation; six samples can be tested at the same time, and optimal temperature can be selected according to the difference between reagent boiling point and ambient temperature so as to achieve quick analysis; reagents can also be recycled to reduce test cost; and soaking, extraction and solvent recovery can be done in one step. Therefore, this device is characterized by reasonable design, stable performance, good reproducibility, high accuracy, easy operation, saving time and effort, and so on.
Scope of application:
SOX406 Fat Analyzer can quickly separate one substance from solid or semi-solid mixtures, can determine the soluble organic compounds contained in foods, feeds, medicines, soil, sludge, polymers, fiber products, petrochemical products, detergents, rubber, plastics and other materials.
Characteristics:
· Integral metal heating, while scope and high precision of temperature control
· Electric circuit is isolated from the extraction space, ensuring devices security
· Over-temperature alarming and timer reminding functions are available
· Timer circuit is isolated from the extraction space, ensuring device security
· Triple alarms i.e. sound, light, LCD screen word prompts are available
· Abundant interface contents give simultaneous display of given temperature, actual temperature, given time and heating time
· The lifting connection of linear bearing conduction technique gives smooth and comfortable lifting operation
· Intelligent man-machine dialogue control system
· Exclusive air insulation technique leaves the case in room temperature, has thermal insulation and temperature maintenance two functions
· 4.3″ LCD screen and microcomputer control system are adopted
Technical data:
Temperature range | Room temperature +5℃~280℃ |
Measuring range | 0.1~100% |
Temperature accuracy | ±1℃ |
Sample weight | 0.5~15g(generally 2~5g,depending on sample) |
Capacity per batch | 6pcs/batch |
Solvent cup volume | 80mL |
Solvent recovery | ≥80% |
Shortened extraction time | 20~80% |
Power supply | 220VAC±10%, 50Hz |
Power | 1KW |
Dimensions | 650mm×320mm×715mm |
Net Weight | 35Kg |
#WD03 Waste gas collection hood
Feature:
#S402 exhausting system
Description
S402 Exhausting System
#SH220F Kjeldahl Digestion System
Description
SH220F Graphite Digester includes globally advanced technology, features quick digestion, high efficiency and easy for use, etc., It’ s widely used in such fields as food, medicine, agriculture, forestry, environmental protection, chemical engineering, biochemistry, as well as universities, research departments and so on, for sample digestion prior to the chemical analysis of soil, feed, plants, seeds, minerals etc., suitable for matching K9840 analyzer.
Characteristic
The SH220F has two temperature control modes: curve heating and linear temperature heating. It can edit and store 10 groups of digestion programs. Each group can set up a segment temperature and time gradient of up to 5 to meet different experimental needs of users.
SH220F
Graphite block have longer life after special anti-oxidation processing.
It adopt advanced PID temperature control technology, high accuracy heating up to 400℃ only cost 20minutes.
WD03 Gas Collection Hood
Using PFA sealed cap, longer life, dealing effect well.
Clip-on sealing cap, easy for changing.
Specialized water jet vacuum pump is adopted, don’t need electric power.
Drip tray design, reduce corrosion damage from acid solutions.
Specification
Model | SH420F | SH220F |
---|---|---|
Temperature range | Room temperature +5 – 450℃ | · |
Temperature accuracy | ±1℃ | · |
Heating method | Infrared heating and high-purity graphite conduction | · |
Heating insulation method | unique air duct insulation technology | · |
Digestion tube capacity | 300mL | · |
Capacity per batch | 20pcs/batch | · |
Power supply | 220 VAC±10% 50HZ | · |
Power | 3600W | · |
Dimensions | 515mmX458mmX730mm | 515mmX421mmX211mm |
Net weight | 40Kg | 25Kg |
#F800 Fiber Analyzer
Feature:
1. hidden solution barrel pull structure designed to facilitate dosing operation, provide the most safety fiber analysis.
2.Provide five various specifications crucible specifications to meet the needs of different samples for standard configuration.
3. the crucible recoil function designed to prevent sample in the crucible cannot caking filtration.
Description
Hanon F800 Fiber Analyzer is with advanced design, easy operation and flexible application. It can be used in conventional Weende method to analyze crude fiber and Van Soest analysis to wash the fiber. Applies to plant, feed, food and other agricultural products as well as the determination crude fiber, neutral detergent fiber (NDF), acid detergent fiber (ADF), hemicellulose and acid detergent lignin (ADL).
Characteristic
1.Hidden solution barrel pull structure designed to facilitate dosing operation, provide the safest fiber analysis.
2.Corrosive liquid is not in contact with any pump body, to avoid waste discharge pump susceptible to corrosion phenomena.
3.The crucible recoil function designed to prevent sample in the crucible can not caking filtration.
4.With dosing overflow protection function to prevent dosing corrosive liquid overflow due to operator error, protect the safety of the operator.
5.Adjust the crucible heating power in time, enabling customers to control the heating rate easily and low down the energy consumption, ECO-friendly.
6.Having a built-in pre-heating function, greatly reducing the whole experiment time.
7.Provide five various specifications crucible specifications to meet the needs of different samples for standard configuration.
8.Can detect crude fiber, neutral detergent fiber (NDF), acid detergent fiber (ADF), hemicelluloses and acid detergent lignin (ADL).
Precise control experiment
Test time can be set free, clockwise and countdown timing functions are available, real-time reminder end of the experiment, the experimenter to facilitate precise control experiment, save test time, improve efficiency.
Integrated infrared heating technology
Advanced integrated infrared heating, more uniform heating of the crucible fast, more consistent sample extraction, higher extraction recovery, thereby improving the accuracy of test results.
Optional peripheral accessories: cool extractor .Experiments can go fat, washed with acetone extraction after detection of acid lignin and other steps.
Specification
Measurement range | 0.1%~100% |
Sample weight | 0.5g~3g |
Repeatability error | Craw Fiber Content below 10%, ≤0.4% | Craw Fiber Content above 10%, ≤1% |
Capacity | 6 pcs/batch |
Pre-heating time | 10-12min |
Heating to boiling | 13-15min |
Rated power | 2.2KW |
Power supply | 220 VAC ±10% 50Hz |
Dimension | 776mm×476mm×644mm |
Visit us | www.elitetradebd.com |
Supplier | Elite Scientific and Meditech CO. |
#Cooking oil tester#OS270 cooking oil tester#OS2700 cooking oil tester
Feature:
Description
Cooking oil is the mainly energy and nutrition source, and the cooking material for daily life. Fried food is also favored as its unique taste. With physical and chemical reaction during the high-temp cooking process, toxical chemicals will be produced. Currently, the traditional testing method on cooking oil is mainly chemical method which is commonly used and conducted by professionals in the lab. But the laboratory process always produces pollution and unnecessary waste. The Total Polar Material (TPM) values is the only acceptable standard which used for cooking oil testing. OS270 is designed for cooking oil testing with safety and efficiency. It is able to test total polar material (TPM) values in minutes and suitable for all kinds of cooking oil within high operation temperature.
OS270/OS2700 Cooking Oil Tester is especially suitable for quality supervision, refectory, fast-food restaurants, food manufacturers, bakeries and so on. It is also used for cooking oil quality’s preliminary inspection in lab. This tester aims to achieve the application and inspecting on oil quality, save cooking oil, and ensure the safety of cooking oil.
Specification
Temperature measure range | 0~200.0 ℃ |
Temperature resolution | ± 0.1℃ |
Temperature accuracy | ± 1.0℃ |
TPM measure range | 0.5~40% |
TPM accuracy | ± 2% (40.0~190.0 ℃) |
TPM resolution | ± 0.1% |
Temperature sensor | PTC Sensor |
TPM sensor | Capacitive sensor |
TPM response time | <15S |
LED status alert | Green,Orange , Red |
Operating temperature | 0 ~+50 ℃ |
Storage temperature | -20 ~+70 ℃ |
Display | LED |
Frame material | ABS |
Probe material | Food-grade stainless steel |
Protection grade | IP 65 |
Battery life below 20 ℃ | About 25 hours of continuous use time (the corresponding measuring 500 times) |
Weight(with the probe cases) | About 165g |
Bluetooth | Yes |
WiFi | Yes |
Auto-Calibrate | Yes |
#Kjeldahl Protein Analyzer#Automatic Nitrogen Analyzer#K1160 Automatic Kjeldahl Protein/K1160 Nitrogen Analyzer
Description
K1160 Automatic Kjeldahl Protein/Nitrogen Analyzer is unique in its intelligence, efficiency, energy saving and data sharing, bringing more new experiences to users. The product can be widely used in the fields of food processing, feed production, tobacco, animal husbandry, soil fertility, environmental monitoring, medicine, agriculture, scientific research, teaching, quality supervision and other fields of nitrogen or protein determination.
Features:
Specifications:
Benefits:
Specification
K1160 Kjeldahl Protein Analyzer | |
Measuring range | 0.1mg ~240mg N |
Analysis time | 3~8min/sample |
Reproducibility | Average value relative error ±0.5% |
Recovery | ≧99.5% |
Burette volume | 1.0μL/step optional:0.2μL/step and 0.4μL/step |
Sample capacity | solid≦5g/sample, liquid≦20mL/sample |
Water consumption in the distillation process | 0.5L/min |
Data storage capacity | 1 million groups |
Power supply | 220VAC±10%, 50Hz |
Power | 2Kw |
Net weight | 38Kg |
Dimensions | 460mm×360mm×725mm |
K1124 Auto Sampler | |
Capacity | 24 samples per batch |
Rotor speed | ≥0.04r/s |
Sampling speed | ≥40mm/s |
Tank capacity | 4×15L |
Connection | CAN |
Voltage | 24V DC |
Rated Power | 40W |
Dimension(L×W×H) | 920mm×625mm×908mm |
Weight | 80Kg |
Visit us | www.elitetradebd.com |
Supplier | Elite Scientific & Meditech CO |
#P810 auto digital Polarimeter
Features
Peltier Temperature Control
The Peltier system automatically controls the sample temperature in the max range from 10 °C to 50 °C.
User Calibration System
Additional Standard polarization tubes with software for calibration by user.
Large Color Touch Screen
Large color touch screen display and innovative software interface give extremely convenient device operation and data acquisition. P810/P850 adopt a 5.6inch TFT display, P810 Pro/ P850 Pro adopt 8inch TFT display.
LED Light Source
High brightness LED lights with service life exceeding 100000 hours. To void the light source exchanging.
Description
P Series Polarimeter with automatic photoelectric inspection technology, boasts accurate and reliable measurement,convenient operation. By detecting the optical rotation, thedensity, content and purity of substances can be analyzed and determined. It’ s widely used in pharmaceutical,petroleum, food, chemicals, flavors, spices, sugar marking and other industries , and universities and research institutes.
New upgraded models: P810 Pro/ P850 Pro adopt brilliant Android system and communication ports, let new models meet FDA CFR21 part 11.
New P810 Pro/ P850 Pro New Characteristics
·Adopt a 8 inch TFT touch screen, with the brilliant Android system, let the operating easier.
·Meet FDA CFR21 part11.
·User ID management function. Users ID are available for adding, deleting and jurisdiction setting.
·Timing measuring and atlas display function, displaying the samples optical rotation changing dynamic process.
·8GB room for data storage.
·Classify and synchronous result searching function is convenience for user’s result searching.
·3 USD port for U-Disk, keyboard&mouse connecting. And WiFi for connecting printer. Let operating much easier.
Specification
Measuring range | ±89.99°Arc;±259°Z |
Measuring Model | Optical Rotation;Specific Rotation;Concentration;SUgar Degree |
Min.reading | 0.001°(optical rotation) |
Accuracy | ±0.01º (-45º≤ optical rotation ≤ +45º) ±0.02º ( optical rotation<﹣45º or optical rotation>﹢45º) |
Repeatability (standard deviation) | 0.002°(optical rotation) |
min.transmittance | 1% |
Wavelength/b> | 589.3nm (sodium D spectrum) |
Temperature Control | No |
Temperature Control Range/b> | |
Temperature Control Accuracy | |
Communication Port | USB;RS232 |
Data Storable | 1000 |
21CFR Part 11 | No |
Display | 5.6inch |
Sample Cell | Upto 200mm |
Atlas Duraltion | No |
Power | 220V AC (±10%)50Hz |
Calibration System | Option |
#SOX606 fat analyzer
Features & Advantages
Descriptions
SOX606 Automatic Soxhlet Extractor, designed based on the Soxhlet extraction principle with weight method to determine the fat content. It has five extraction methods to meet different demands from customer. The whole metal heating module, heating up fast, good effect, low power consumption; access to all-round water temperature, flow control, environmental protection more economical; built-in ether leak detection device to ensure experimental safety. It is widely used in agricultural/ food /chemical /environment and other areas, especially for the drug ,soil, sludge,cleaner and other substances in the extraction of soluble organic compounds.
Automatic extraction
One touch to start extraction, lifting, pre-heating, extraction, dilution, solvent recovery automatically.
In-built 5 unique methods of extraction, handy operation.
Include Soxhlet Extraction, Soxhlet Hot Extraction, Continue Flow Extractioin, Hot Extraction, CH Standard methods,
Apply to all organic solvents
Customized by glass and PTFE, which can block all impurities and possess high rate of accuracy, apply to all organic solvents, satisfying the requirements of users.
ECO- friendly and conservation
Full temperature and flow control for reduce consumption of condensed water.
Specifications
Technical data:
Temperature range | Room temperature +5℃~300℃ |
Measuring range | 0.1~100% |
Temperature accuracy | ±1℃ |
Reproducibility | Relative error 1% |
Sample weight | 0.5~15g |
Capacity per batch | 6pcs/batch |
Solvent cup volume | 150mL |
Solvent recovery | ≧85% |
Power supply | 220VAC±10%, 50Hz |
Power | 2.6KW |
Dimensions | 650mm×380mm×720mm |
Net Weight | 50Kg |
#SH420F Kjeldahl Digestion System
SH420F Kjeldahl Digestion adopts globally advanced high-temperature infrared t radiation heating technology and microprocessor control platform, boasts accurate temperature control and quick temperature rise. It has two kinds of temperature rise mode: linear and curve temperature rise mode, and offers 20 digestion programs for control of temperature rise curves.
Neutralization system has many functions such as triple filtration, condensate recovery of exhaust gas, filtration and neutralization device. The product adopts high-quality Anticorrosive Pumps, low noise, strong suction, reduce exhaust emissions, co-friendly.
Description
S402 Exhausting System
Specification
Model | SH420F | SH220F |
Temperature range | Room temperature +5 – 450℃ | |
Temperature accuracy | ±1℃ | · |
Heating method | Infrared heating and high-purity graphite conduction | · |
Heating insulation method | unique air duct insulation technology | · |
Digestion tube capacity | 300mL | · |
Capacity per batch | 20pcs/batch | · |
Power supply | 220 VAC±10% 50HZ | · |
Power | 3600W | · |
Dimensions | 515mmX458mmX730mm | 515mmX421mmX211mm |
Net weight | 40Kg | 25Kg |