Elite Scientific & Meditech CO.

ICU ventilator

0.00৳ 

Application:

The ventilator is suitable for various kinds of medical institutions for cardiopulmonary resuscitation respiratory support. Acute respiratory failure caused by various reasons or incomplete oxygenation dysfunction. Intra-operation,  post-operation respiratory support, other respiratory treatment.

Trust Points

l Simplicity: easy to use, easy to move with 4 wheels.

l Choice: adapt the equipment to your patients and procedures freely

l Patient Centered Ventilation: Precision in an anesthesia ventilator, from conventional ventilation to advanced modes, including 8 modes: IPPV; A/C; PCV; SIMV; SPONT/CPAP;PSV;SIGH;MANU.

l Designed and manufactured by Perlong Medical with over 23 years experience in this area.

l Flexible configurations to suit your needs.

l International standard and advanced technology suitable for wide range use.

l Compact interface and big screen give you better operating experience.

l Over 2,000 units installed in the world.

Features

l 12.1” TFT LCD screen displays the Ventilation parameters, Alarming information and Waveform.

l Recycle breathing pipe, ensure easy operating and keep tidy.

l Multiple working modes such as volume control and pressure limit, adapt to wide range patient.

l Vaporizer with temperature, flow compensation and self-lock function, keep safety anytime.

l Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects;

l Real time pressure-time, flow-time loop graphics and high precision O2 concentration detection function included.

l Stable and low noise air compressor, create quiet work environment for doctors.

Safety

l Three level alarming system, visual and sound alarm information.

l With lots of alarming, reminding and protection functions.

l Advanced power management control technology.

l With built-in backup power source, when outside power source goes off, back-up power source starts to work.

l Self-check before running, eliminate system mistake.

l Separate design of electric and gas, keep safety running of ventilator.

Specifications

Ventilation modes: IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV,SIGH,MANU
Ventilator parameter ranges
Tidal volume(Vt) 0 ~ 2000mL
Frequency (Freq) 

Oxygen concentration

1bpm ~ 100 bpm 

21%-100%

I/E 4:1~1:8
PEEP 0cmH2O ~ 40 cmH2O
Pressure Limit 20 cmH2O ~ 100 cmH2O
Monitored Parameters
Frequency (Freq) 0 /min ~ 100 /min
Tidal volume(Vt) 0 mL ~ 2500 mL
Oxygen concentration 15 % ~ 100 %
Oscillographs display:
P-T(pressure – time)
F-T(flow – time)
P-V loop (pressure – volume loop)
Size 

1st Wooden case packing size(main engine):L560*W 560*H 605mm , GW:35KG ; NW:17KG

2nd Wooden case packing size(air compressor):L670*W700*H 1160 mm , GW:79KG ; NW:46.2KG

Alarm and protection

The AC power failure alarm Power failure or no connection
Internal battery backup low voltage alarm ≤11.3±0.3V
No tidal volume No tidal volume within 6s
High Minute Volume alarm 

LowMinute Volume alarm

5L/min-99L/min 

1L/min-30L/min

High Airway pressure alarm 

Low Airway pressure alarm

High oxygen concentration alarm

Low oxygen concentration alarm

20cmH2O-100cmH2O 

0cmH2O-20cmH2O

19%-100%

18%-99%

Continuous pressure alarm (PEEP+1.5cmH2O) over 16s
Suffocation warning 

Fan error

Oxygen deficit

5-60s 

Show on screen

Show on screen

The maximum limited pressure <12.5 kPa
Working conditions
Gas source O2,AIR
Pressure 280kPa-600kPa
Voltage -220V±22V
Power frequency 50Hz±1Hz
Input power 900VA(with air compressor) 

250VA(without air compressor)

Scroll to Top